Product Number: ELK23J0515
Manual: Download PDF

Plant SE ELISA Kit instruction

Intended use

This SE ELISA kit is intended Laboratory for Research use only and is not for use in diagnostic or

therapeutic procedures.The Stop Solution changes the color from blue to yellow and the intensity of the

color is measured at 450 nm using a spectrophotometer. In order to measure the concentration of SE in the

sample, this SE ELISA Kit includes a set of calibration standards. The calibration standards are assayed at

the same time as the samples and allow the operator to produce a standard curve of Optical Density versus

SE concentration. The concentration of SE in the samples is then determined by comparing the O.D. of the

samples to the standard curve.

Sample collection and storages

Serum - Use a serum separator tube and allow samples to clot for 30 minutes before centrifugation for

10 minutes at approximately 3000×g. Remove serum and assay immediately or aliquot and store samples

at -20 ℃ or -80 ℃ .Avoid repeated freeze-thaw cycles.

Plasma - Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge samples for 30 minutes

at 3000×g at 2-8 ℃ within 30 minutes of collection. Store samples at -20 ℃ or -80 ℃ . Avoid repeated

freeze-thaw cycles.

Cell culture supernates and other biological fluids - Remove particulates by centrifugation and

assay immediately or aliquot and store samples at -20 ℃ or -80 ℃ . Avoid repeated freeze-thaw cycles.

Note:

The samples shoule be centrifugated dequately and no hemolysis or granule was

allowed.

Materials required but not supplied

1.

Standard microplate reader(450nm).

2.

Precision pipettes and Disposable pipette tips.

3.

37 ℃ incubator.

Precautions

1.

Do not substitute reagents from one kit to another. Standard, conjugate and microplates are

matched for optimal performance. Use only the reagents supplied by manufacturer.

2.

Do not remove microplate from the storage bag until needed. Unused strips should be stored at

2-8°C in their pouch with the desiccant provided.

3.

Mix all reagents before using.

Remove all kit reagents from refrigerator and allow them to reach room temperature ( 20-25°C).

Materials supplied

Name 96 determinations 48 determinations
Microelisa stripplate 8*12strips 8*6strips
Standard 0.3ml*6tubes 0.3ml*6tubes
Sample Diluent 6.0ml 3.0ml
HRP-Conjugate reagent 10.0ml 5.0ml
20X Wash solution 25ml 15ml
Chromogen Solution A 6.0ml 3.0ml
Chromogen Solution B 6.0ml 3.0ml
Stop Solution 6.0ml 3.0ml
Closure plate membrane 2 2
User manual 1 1

Reagent preparation

20×wash solution:Dilute with Distilled or deionized water 1:20.

Assay procedure

1.

Prepare all reagents before starting assay procedure. It is recommended that all Standards and

Samples be added in duplicate to the Microelisa Stripplate.

2.

Add standard: Set Standard wells, testing sample wells. Add standard 50μl to standard well.

3.

Add Sample: Add testing sample 10μl then add 40μl of Sample Diluent to testing sample wel.

4.

Add 100μl of HRP-conjugate reagent to each well, c over with an adhesive strip and incubate for 60

minutes at 37°C.

5.

Aspirate each well and wash, repeating the process four times for a total of five washes. Wash by

filling each well with Wash Solution (400 μl ) using a squirt bottle, manifold dispenser or autowasher.

Complete removal of liquid at each step is essential to good performance. After the last wash, remove

any remaining Wash Solution by aspirating or decanting. Invert the plate and blot it against clean paper

towels.

6.

Add chromogen solution A 50μl and chromogen solution B 50μl to each well. Gently mix and

incubate for 15 minutes at 37°C. Protect from light.

7.

Add 50 μl Stop Solution to each well. The color in the wells should change from blue to yellow. If the

color in the wells is green or the color change does not appear uniform, gently tap the plate to ensure

thorough mixing.

8

Read the Optical Density (O.D.) at 450 nm using a microtiter plate reader within 15 minutes.

Calculation of results

1. This standard curve is used to determine the amount in an unknown sample. The standard curve is

generated by plotting the average O.D. (450 nm) obtained for each of the six standard concentrations on

the vertical (Y) axis versus the corresponding concentration on the horizontal (X) axis.

2. First, calculate the mean O.D. value for each standard and sample. All O.D. values, are subtracted by

the mean value of the zero standard before result interpretation. Construct the standard curve using graph

paper or statistical software.

3. To determine the amount in each sample, first locate the O.D. value on the Y-axis and extend a

horizontal line to the standard curve. At the point of intersection, draw a vertical line to the X-axis and

read the corresponding concentration.

4. Any variation in operator, pipetting and washing technique, incubation time or temperature, and kit age

can cause variation in result. Each user should obtain their own standard curve.

5. The sensitivity by this assay is 1 ng/mL.

6. Standard curve.

Storage :

2-8 ℃ .

Validity : six months.

FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC

APPLICATIONS! PLEASE READ THROUGH ENTIRE PROCEDURE BEFORE

BEGINNING!