User Manual
ELK17J0862, Mouse Urolithin A
standards concentration (X)
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
Mouse Urolithin A ELISA Kit instruction
Intended use
This Urolithin A ELISA kit is intended Laboratory for Research use only
and is not for use in diagnostic
or therapeutic procedures.The Stop Solution changes the color from blue
to yellow and the intensity of the
color is measured at 450 nm using a spectrophotometer. In order to
measure the concentration of Urolithin
A in the sample, this Urolithin A ELISA Kit includes a set of
calibration standards. The calibration
standards are assayed at the same time as the samples and allow the
operator to produce a standard curve
of Optical Density versus Urolithin A concentration. The concentration
of Urolithin A in the samples is
then determined by comparing the O.D. of the samples to the standard
curve.
Sample collection and storages
Serum - Use a serum separator tube and allow samples to clot for 30
minutes before centrifugation for
10 minutes at approximately 3000×g. Remove serum and assay immediately
or aliquot and store samples
at -20℃or -80℃.Avoid repeated freeze-thaw cycles.
Plasma - Collect plasma using EDTA or heparin as an anticoagulant.
Centrifuge samples for 30 minutes
at 3000×g at 2-8℃within 30 minutes of collection. Store samples at
-20℃or -80℃. Avoid repeated
freeze-thaw cycles.
Cell culture supernates and other biological fluids - Remove
particulates by centrifugation and
assay immediately or aliquot and store samples at -20℃or -80℃. Avoid
repeated freeze-thaw cycles.
Note: The samples shoule be centrifugated dequately and no hemolysis or
granule was
allowed.
Materials required but not supplied
| 6.0ml | 3.0ml |
|---|---|
| 6.0ml 6.0ml | 3.0ml 3.0ml |
| membrane 2 | 2 |
| 1 Standard (S0 →S5) concentration was followed by:0,10,20,40,80,160 ng/mL. | 1 |
| 20×wash solution:Dilute with Distilled or deionized water 1:20. | |
| Prepare all reagents before starting assay procedure. It is recommended that all Standards and | |
| Add standard: Set Standard wells, testing sample wells. Add standard 50μl to standard well. | |
| Add Sample: Add testing sample 10μl then add 40μl of Sample Diluent to testing sample wel. | |
| any remaining Wash Solution by aspirating or decanting. Invert the plate and blot it against clean paper towels. |