User Manual
ELK05J0268, Escherichia coli host cell protein (E.coli P)
| --- | --- | --- | --- |
LOGIT- LOG
standards concentration (X)
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
E.coli P ELISA Kit instruction
Intended use
This E.coli P ELISA kit is intended Laboratory for Research use only and
is not for use in diagnostic or therapeutic procedures.The Stop Solution
changes the color from blue to yellow and the intensity of the color is
measured at 450 nm using a spectrophotometer. In order to measure the
concentration of E.coli P in the sample, this E.coli P ELISA Kit
includes a set of calibration standards. The calibration standards are
assayed at the same time as the samples and allow the operator to
produce a standard curve of Optical Density versus E.coli P
concentration. The concentration of E.coli P in the samples is then
determined by comparing the O.D. of the samples to the standard
curve.
Sample collection and storages
- Can’t detect the samples which contain NaN3, because NaN3 inhibits
HRP activity of the horseradish peroxidase.
- Extract as soon as possible after Specimen collection, Extracted
according to the relevant literature. Cell culture supernates and plant
exact fluids - Remove particulates by centrifugation and assay
immediately or aliquot and store samples at -20°C or -80°C. Avoid
repeated freeze-thaw.
Materials required but not supplied
- Standard microplate reader(450nm)
- Precision pipettes and Disposable pipette tips.
- 37 ℃incubator
Precautions
- Do not substitute reagents from one kit to another. Standard,
conjugate and microplates are matched for optimal performance. Use only
the reagents supplied by manufacturer.
- Do not remove microplate from the storage bag until needed. Unused
strips should be stored at
2-8°C in their pouch with the desiccant provided.
- Mix all reagents before using.
Remove all kit reagents from refrigerator and allow them to reach room
temperature ( 20-25°C)
Materials supplied
Name 96 determinations 48 determinations
Microelisa stripplate 8*12strips 8*6strips
Standard 0.3ml*6tubes 0.3ml*6tubes
Sample Diluent 6.0ml 3.0ml
HRP-Conjugate reagent 6.0ml 3.0ml
20X Wash solution 25ml 15ml
Chromogen Solution A 6.0ml 3.0ml
Chromogen Solution B 6.0ml 3.0ml
Stop Solution 6.0ml 3.0ml
Closure plate
membrane 2 2
User manual 1 1
Sealed bags 1 1
Note: Standard (S0 →S5) concentration was followed by:0、5、10、20、40、80
ng/mL
Reagent preparation
20×wash solution:Dilute with Distilled or deionized water 1:20.
Assay procedure
- Prepare all reagents before starting assay procedure. It is
recommended that all Standards and Samples be added in duplicate to the
Microelisa Stripplate.
- Add standard: Set Standard wells, testing sample wells. Add standard
50μl to standard well.
- Add Sample: Add testing sample 10μl then add Sample Diluent 40μl to
testing sample well;
- Add 50μl of HRP-conjugate reagent to each well, cover with an
adhesive strip and incubate for 60 minutes at 37°C.
- Aspirate each well and wash, repeating the process four times for a
total of five washes. Wash by filling each well with Wash Solution
(400μl) using a squirt bottle, manifold dispenser or autowasher.
Complete removal of liquid at each step is essential to good
performance. After the last wash, remove
any remaining Wash Solution by aspirating or decanting. Invert the plate
and blot it against clean paper
towels.
- Add chromogen solution A 50μl and chromogen solution B 50μl to each
well. Gently mix and
incubate for 15 minutes at 37°C. Protect from light.
- Add 50μl Stop Solution to each well. The color in the wells should
change from blue to yellow. If the
color in the wells is green or the color change does not
appear uniform, gently tap the plate to ensure thorough mixing.
- Read the Optical Density (O.D.) at 450 nm using a microtiter plate
reader within 15 minutes.
Calculation of results
- Calculate the mean absorbance value A450 for each set of reference
standards and samples.
- Divide the average A450 value for each standard, control and test
sample by the average A450 of
standard0 and multiply by 100 to obtain %B/B0 for each sample.
- Prepare a standard curve by plotting the average absorbance of each
standard versus the
corresponding concentrations of the standards on linear-log graph paper
or the %B/B0 value for each
standard versus the corresponding concentration of the standard on
linear-log or logit-log graph paper.
logit=ln(B/ B0)/(1-B/ B0)
- Any values obtained for diluted samples must be further converted by
applying the appropriate
dilution factor in the calculations.
- The standard density is a X, the B/ B0 is a Y, sitting to mark the
paper in the logit- log up draw a
standard curve. According to the B/ B0 that need to be measured the
sample can from sit to mark the
density value that the paper looks up the sample up.
- The sensitivity by this assay is 1ng/mL
- The standard curve presented here is an example of the data
typically produced with this kit;
however, your results will not be identical to these.
- Standard curve match
LOGIT- LOG
B/()
standards concentration (X)
Storage: 2-8℃. Validity:six months.
FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS!
PLEASE READ THROUGH ENTIRE PROCEDURE BEFORE BEGINNING!