MEBEP TECH(HK) Co., Limited
Email: sales@mebep.com Website: http://www.mebep.com
Tel: +86-755-86134126 WhatsApp/Facebook/Twitter: +86-189-22896756
User Manual
ELK02J0625, Rat Pro-interleukin-1β (Pro-IL-1β)
- Do not substitute reagents from one kit to another. Standard,
conjugate and microplates are matched for optimal performance. Use only
the reagents supplied by manufacturer.
- Do not remove microplate from the storage bag until needed. Unused
strips should be stored at 2-8°C in their pouch with the desiccant
provided.
- Mix all reagents before using.
Remove all kit reagents from refrigerator and allow them to reach room
temperature ( 20-25°C).
Materials supplied
Name 96 determinations 48 determinations
Microelisa stripplate 8*12strips 8*6strips
Standard 0.3ml*6tubes 0.3ml*6tubes
Sample Diluent 6.0ml 3.0ml
HRP-Conjugate reagent 10.0ml 5.0ml
20X Wash solution 25ml 15ml
Chromogen Solution A 6.0ml 3.0ml
Chromogen Solution B 6.0ml 3.0ml
Stop Solution 6.0ml 3.0ml
Closure plate
membrane 2 2
User manual 1 1
Note: Standard (S0 →S5) concentration was followed by:0,5,10,20,40,80
pg/mL.
Reagent preparation
20×wash solution:Dilute with Distilled or deionized water 1:20.
Assay procedure
- Prepare all reagents before starting assay procedure. It is
recommended that all Standards and Samples be added in duplicate to the
Microelisa Stripplate.
- Add standard: Set Standard wells, testing sample wells. Add standard
50μl to standard well.
- Add Sample: Add testing sample 10μl then add 40μl of Sample Diluent
to testing sample wel.
- Add 100μl of HRP-conjugate reagent to each well, cover with an
adhesive strip and incubate for 60 minutes at 37°C.
- Aspirate each well and wash, repeating the process four times for a
total of five washes. Wash by filling each well with Wash Solution
(400μl) using a squirt bottle, manifold dispenser or autowasher.
Complete removal of liquid at each step is essential to good
performance. After the last wash, remove
any remaining Wash Solution by aspirating or decanting. Invert the plate
and blot it against clean paper towels.
- Add chromogen solution A 50μl and chromogen solution B 50μl to each
well. Gently mix and incubate for 15 minutes at 37°C. Protect from
light.
- Add 50μl Stop Solution to each well. The color in the wells should
change from blue to yellow. If the color in the wells is green or the
color change does not appear uniform, gently tap the plate to ensure
thorough mixing.
8 Read the Optical Density (O.D.) at 450 nm using a microtiter plate
reader within 15 minutes.
Calculation of results
- This standard curve is used to determine the amount in an unknown
sample. The standard curve is generated by plotting the average O.D.
(450 nm) obtained for each of the six standard concentrations on the
vertical (Y) axis versus the corresponding concentration on the
horizontal (X) axis.
- First, calculate the mean O.D. value for each standard and sample.
All O.D. values, are subtracted by the mean value of the zero standard
before result interpretation. Construct the standard curve using graph
paper or statistical software.
- To determine the amount in each sample, first locate the O.D. value
on the Y-axis and extend a horizontal line to the standard curve. At the
point of intersection, draw a vertical line to the X-axis and read the
corresponding concentration.
- Any variation in operator, pipetting and washing technique,
incubation time or temperature, and kit age can cause variation in
result. Each user should obtain their own standard curve.
- The sensitivity by this assay is 1 pg/mL.
- Standard curve.
Storage: 2-8℃. Validity:six months.
FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS!
PLEASE READ THROUGH ENTIRE PROCEDURE BEFORE BEGINNING!